[PROVA] PROVA-19/21/23 누설전류계
제품기능 및특징
Active(W, KW, HP), reactive(VAR, KVAR) 측정Apparent(VA, KVA) power 측정
Power factor(PF), Phase Angle
Energy(mWH, WH, KWH)
IT 제품용 standby power consumption 측정
Non-interrupted AC current hamonic 분석
1-99th order of harmonics at 1.0% basic accuracy
Total harmonic distortion, Crest Factor
전압, 전류의 실효치 측정 at 0.5% basic Accuracy
빠른 peak 기능
평형 3상 전력, 평형 3상 검상
CT율 1-250까지 조절 가능
최대값, 최소값, data hold 기능
Active power in H.P
자동전원꺼짐(30분)
-Model 19, 21(외부간섭에 대한 클램프 보호기능)
제품사양
AC Watt (50 or 60 ㎐, PF 0.6 to 1, CT=1, V는 AC 4A보다 높을 때, 전류는 mA 범위에서
AC 1mA보다
높을 때, A범위에서는 AC 0.04A보다 높을 때임. Specification은 연속 waveform에
적용됨)
Model 21
|
측정범위(0-30A) |
분해능 |
Accuracy of Readings1,2 |
|
0.050 - 9.999W |
0.001W |
±2%±0.025W |
|
10.00 - 99.99W |
0.01W |
±2%±0.25W |
|
100.0 - 999.9W |
0.1W |
±2%±2.5W |
|
1.000 - 9.999㎾ |
0.001㎾ |
±2%±0.025KW |
|
10.00 - 99.99㎾ |
0.01㎾ |
±2%±0.25KW |
|
100.0 - 999.9㎾ |
0.1㎾ |
±2%±2.5KW |
|
1000 - 9999㎾ |
1㎾ |
±2%±25W |
Model 23
|
측정범위(0-60A) |
분해능 |
Accuracy of Readings1,2 |
|
0.050 - 9.999W |
0.001W |
±2%±0.025W |
|
10.00 - 99.99W |
0.01W |
±2%±0.25W |
|
100.0 - 999.9W |
0.1W |
±2%±2.5W |
|
1.000 - 9.999㎾ |
0.001㎾ |
±2%±0.025KW |
|
10.00 - 99.99㎾ |
0.01㎾ |
±2%±0.25KW |
|
100.0 - 999.9㎾ |
0.1㎾ |
±2%±2.5KW |
|
1000 - 9999㎾ |
1㎾ |
±2%±25W |
※ 1: CT≠1이면, 퍼센트의 정확도는 같다(±2%). 그러나 추가적인
와트량은 CT율에 의해 곱해져야 한다.
예를 들면, ±0.025W는 0.025W*CT율이 된다.
※ 2: 만약 자동 Hz가 선택되면, AC V는 50V보다 커야하며 2˚
위상각error를 정확도에 더한다.
|
Model 21 범위(30-50A)
Model 23 범위(60-100A) |
분해능 |
Accuracy of Readings3,4 |
|
0.050 - 9.999W |
0.001W |
±2%ofVA±5dgts |
|
10.00 - 99.99W |
0.01W |
±2%ofVA±5dgts |
|
100.0 - 999.9W |
0.1W |
±2%ofVA±5dgts |
|
1.000 - 9.999㎾ |
0.001㎾ |
±2%ofVA±5dgts |
|
10.00 - 99.99㎾ |
0.01㎾ |
±2%ofVA±5dgts |
|
100.0 - 999.9㎾ |
0.1㎾ |
±2%ofVA±5dgts |
|
1000 - 9999㎾ |
1㎾ |
±2%ofVA±5dgts |
※ 3: CT≠1이면, 퍼센트의 정확도는
같다(±2%). 그러나 추가적인 digit은 CT율에 의해 곱해져야 한다.
예를 들면, ±5 digits는 ±5 digits*CT율이 된다.
※ 4: 만약 자동 Hz가 선택되면, AC
V는 50V보다 커야 한다.
AC Current (50 or 60 ㎐, Auto Range, True RMS, Crest Factor<4,
CT=1, Overload Protection AC 600A)
Model 19, 21
|
측정범위 |
분해능 |
Accuracy of Readings5,6 |
|
0.50 - 60.00mA |
0.01mA |
±0.5%±5dgts |
|
60.00 - 99.99mA |
0.01mA |
±0.5%±50dgts |
|
100.0 - 600.0mA |
0.1mA |
±0.5%±5dgts |
|
0.050 - 3.000A |
0.001A |
±0.5%±5dgts |
|
3.000 - 9.99A |
0.001A |
±0.5%±50dgts |
|
10.00 - 30.00A |
0.01A |
±0.5%±5dgts |
|
30.00 - 50.00A |
0.01A |
±1.0%±5dgts |
※ 5: CT≠1이면, 퍼센트의 정확도는
같다(±0.5%). 그러나 추가적인 digit은 CT율에 의해 곱해져야 한다.
예를 들면, ±5 digits는 ±5 digits*CT율이 된다.
※ 6: 만약 자동 Hz가 선택되면, AC
Current는 mA 범위에서 10mA보다 커야 하고 A범위에서는 1A보다 커야 한다.
Model 23
|
측정범위 |
분해능 |
Accuracy of Readings7,8 |
|
0.50 - 60.00mA |
0.01mA |
±0.5%±5dgts |
|
60.00 - 99.99mA |
0.01mA |
±0.5%±50dgts |
|
100.0 - 600.0mA |
0.1mA |
±0.5%±5dgts |
|
0.050 - 9.999A |
0.001A |
±0.5%±5dgts |
|
10.00 - 60.00A |
0.01A |
±0.5%±50dgts |
|
60.00 - 99.99A |
0.01A |
±1.0%±5dgts |
※ 7: CT≠1이면, 퍼센트의 정확도는
같다(±0.5%). 그러나 추가적인 digit은 CT율에 의해 곱해져야 한다.
예를 들면, ±5 digits는 ±5 digits*CT율이 된다.
※ 8: 만약 자동 Hz가 선택되면, AC
Current는 mA 범위에서는 10mA보다 커야 하고 A범위에서는 1A보다 커야 한다.
AC Voltage
(50 or 60㎐, True RMS, Crest Factor<4, Input Impedance 10MΩ,
Overload Protection AC 800V)
Model 19,21,23
|
측정범위 |
분해능 |
Accuracy of Readings9 |
|
5V - 250V |
0.1V |
±0.5±5dgts |
|
250V - 600V |
0.1V |
±0.5±5dgts |
※ 9: 만약 자동 Hz가 선택되면, AC
V는 50V보다 커야한다.
Harmonics of AC Voltage in percentage(1 to 99th order에서 50이나 60Hz일 때
최소 V는 AC 80V보다 커야한다. 만약 V가 50이나 60Hz에서 0이라면 %는 0으로 표시된다)
|
측정범위(Prova-19,21) |
분해능 |
정확도 |
|
1-10th` |
0.1% |
±1% of reading ±1% |
|
11-20th |
0.1% |
±5% of reading ±1% |
|
21-50th |
0.1% |
±15% of reading ±1% |
|
51-99th |
0.1% |
±35% of reading ±1% |
|
측정범위(Prova-23) |
분해능 |
정확도 |
|
1-10th` |
0.1% |
±0.7% of reading ±1% |
|
11-20th |
0.1% |
±2% of reading ±1% |
|
21-50th(A range) |
0.1% |
±5% of reading ±1% |
|
21-50th(mA range) |
0.1% |
±10% of reading ±1% |
|
51-99th |
0.1% |
±35%of reading ±1% |
Harmonics of AC Voltage in Magnitude
(1 to 99th order에서 50이나 60Hz일 때
AC 전류는 80V보다 커야 한다.)
|
측정범위(Prova-19,21) |
분해능 |
정확도 |
|
1-10th` |
0.1V |
±1% of reading ±7dgts |
|
11-20th |
0.1V |
±5% of reading ±7dgts |
|
21-50th |
0.1V |
±15% of reading ±7dgts |
|
51-99th |
0.1V |
±35% of reading ±7dgts |
|
측정범위(Prova-23) |
분해능 |
정확도 |
|
1-10th` |
0.1V |
±0.7% of reading ±7dgts |
|
11-20th |
0.1V |
±2% of reading ±7dgts |
|
21-50th(A range) |
0.1V |
±5% of reading ±7dgts |
|
21-50th(mA range) |
0.1V |
±10% of reading ±7dgts |
|
51-99th |
0.1V |
±35% of reading ±7dgts |
Harmonics of AC Current in Percentage
(1 to 99th order에서 50이나 60Hz일 때
최소 전류는 mA 범위에서 100mA를, A 범위에서는 1A보다 커야 한다.
만약 50, 60Hz에서 전류가 0이라면 Percentage는 0으로 표시된다.)
|
측정범위(Prova-19,21) |
분해능 |
정확도 |
|
1-10th |
0.1% |
±1% of reading ±1% |
|
11-20th |
0.1% |
±5% of reading ±1% |
|
21-50th |
0.1% |
±15% of reading ±1% |
|
51-99th |
0.1% |
±35% of reading ±1% |
|
측정범위(Prova-23) |
분해능 |
정확도 |
|
1-10th` |
0.1% |
±0.7% of reading ±1% |
|
11-20th |
0.1% |
±2% of reading ±1% |
|
21-50th(A range) |
0.1% |
±5% of reading ±1% |
|
21-50th(mA range) |
0.1% |
±10% of reading ±1% |
|
51-99th |
0.1% |
±35% of reading ±1% |
Harmonics of AC Current in% Magnitude(1 to 99th order에서 50이나 60Hz일 때
최소 전류가 100mA보다 커야한다.-mA 범위일 때-, 그리고 A 범위일 때는 1A보다 커야 한다)
Model 19,21,23
|
측정범위 |
분해능 |
정확도 |
|
1-10th` |
0.01mA/0.1mA/
0.001A/0.01A |
reading in mA:±1%of reading±2mA
reading in A:±1%of reading±0.3A |
|
11-20th |
reading in mA:±7%of reading±2mA
reading in A:±7%of reading±0.3A | |
|
21-50th |
reading in mA:±15%of reading±3mA
reading in A:±15%of reading±0.3A | |
|
51-99th |
reading in mA:±35%of reading±3mA
reading in A:±35%of reading±0.3A |
|
Power Factor(PF) (Model 21 and 23) |
Phase Angle(?) (Model 21 and 23) | ||||
|
측정범위 |
분해능 |
정확도 |
측정범위 |
분해능 |
정확도 |
|
0.000 -
1.000 |
0.001 |
±0.04 |
-180° to
180°(0° to 360°) |
0.1° |
±2° |
Total Harmonic Distortion
Model 19,21
|
측정범위 |
분해능 |
정확도 |
|
0.0-10.0% |
0.1% |
±2% |
|
10.0-40% |
0.1% |
5% of reading ±5% |
|
40-100% |
0.1% |
10% of reading ±10% |
|
100-999.9% |
0.1% |
20% of reading |
Model 23
|
측정범위 |
분해능 |
정확도 |
|
0.0-20.0% |
0.1% |
±2% |
|
20.0-100% |
0.1% |
±3% of reading ±5% |
|
100-999.9% |
0.1% |
±10% of reading ±10% |
Peak value of periodic Voltage
Model 19,21,23
|
측정범위 |
Sampling Time |
Accuracy of Reading |
|
50Hz |
39μs |
±5% of 30 digits |
|
60Hz |
33μs |
±5% of 30 digits |
Crest Factor (C.F.,Accuracy of Readings)
|
측정범위 |
분해능 |
정확도 |
|
1.00 - 99.99 |
0.01 |
±5%±30dgts |
Frequency(Hz)
Model 19,21,23
|
측정범위 |
분해능 |
Accuracy of reading |
Allowed Input |
|
mA(45-65Hz) |
0.1Hz |
±0.5Hz |
20mA to 1.2A |
|
A(45-65Hz) |
0.1Hz |
±0.5Hz |
1A to 100A |
제품사양
|
세부사항 |
내 용 |
|
클램프 사이즈 |
30mm(Approx.) |
|
표시 |
4+2+2 Digits LCD |
|
측정범위 선택 |
자동 |
|
Overload 표시 |
OL |
|
전력 소비 |
약 10mA |
|
Low 배터리 표시 |
![]() |
|
자동꺼짐기능 |
약 30분 |
|
기간당 샘플 수 |
512(Voltage or Current) 256(Power) |
|
LCD update rate |
2 times / sec |
|
사용 온도/습도 |
-10°C ~ 50°C / 0 ~ 85% RH |
|
보관 온도/습도 |
-20°C ~ 60°C / 0 ~ 75% RH |
|
전원 |
two 1.5V SUM-3 |
|
크기/무게 |
210mm(L)x62mm(W)x35.6mm(H) / 200g(Battery 포함) |
|
기본악세사리 |
사용설명서, Battery (1.5Vx2EA), Carrying Case, Test
Leads |
|
옵션 |
Alligator Clips |