Altek 322 4-Type Multiple Thermocouple Calibrator
Calibrate All Your T/C
Inputs
Alarms, recorders, controllers and indicators with 0.1°
resolution.
Built-In T/C
Thermometer
Connect the T/C and accurately read the temperature.
Instantly recall the minimum and maximum temperatures at any time to check the
process.
Unparalleled
Accuracy
422
& 322 ±(0.008% of reading + 0.006 mV).
Accurate to ±0.2 °C/0.4 °F for
types J, K, T, E, N, L & U.
Accurate to ±0.7 °C/1.3 °F for types R, S, B,
G, C, D & P.
Millivolt range accurate to ±0.006 mV!
Quik-Chek
Switch
Instantly recall three output temperatures per
type.
Automatic Cold Junction
Compensation
Miniature T/C connector and screw terminals on Models
422 and 322 for stable connections to exclusive Isothermal Block.
Display Directly in °C
& °F
Model 322-1four types J, T, E, K & mV.
Model
322-2four types B, R, S, K & mV.
Optional Wire Kits with
Miniature T/C Connector On One End
Kit-1 types J, K, T, E.
Kit-2
types B, R/S, K.
Source and read T/C’s
over the entire industrial temperature range with Altek’s Model 322 Thermocouple
Calibrator. Use with transmitters, recorders, controllers, alarms, indicators,
data acquisition and computer systems. Switch between four T/C types or
millivolts.
High accuracy and stability is achieved through Altek’s
exclusive isothermal block. Make your connections directly with thermocouple
wire or with miniature thermocouple connectors.
Field customize the Model
322 to lock-in 0.1° or 1° resolution, fixed °F or °C or front panel selectable
°F/°C operation. A shrouded miniature thermocouple connector receptacle plus
terminal screws allow for easy hookups. Built in protection guards the Model 322
against mis-connection to 120 Volts AC or DC, in any mode.
The Model 322
turns on to the T/C type last used. Other T/C types may be selected each time
the unit is turned on. Other T/C types may be selected each time the unit is
turned on. If you always use just one T/C type, lock in the selected T/C type
with the internal DIP switch to prevent accidental change to an unwanted
type.
Select resolution of 0.1° or 1° for the full listed range of all
thermocouple types. Millivolts allows 1 microvolt resolution from -99.999 to
99.999mV. The Model 322 simulates key temperatures for repetitive calibrations.
"Quik-Chek" function stores three output temperatures for each T/C type (15
total) for real convenience. Three memories are retained for each thermocouple
type even when the power is off.
Turn the knob to check trip points,
controller action or hysteresis. The fast response 322 sets quickly without
overshoot but allows slow changes at your rate.
The Model 322 display
gives you fast, accurate temperature measurement with 0.1 and 1 degree or with
0.001 millivolt resolution. High resistance or open T/Cs and leads are detected
and indicated on the LCD display. Two readings per second track fast moving
temperatures.
"MAX" and "MIN" memories are continuously updated from
turn-on or whenever the "RESET" pushbutton is pressed. The Model 322 gives you a
handy tool to monitor temperatures for drift or control deviation. Just flip the
"Quik-Chek" switch to display the minimum and maximum temperature since
reset.
Each time the Model 322 is turned on, the LCD will display all
segments for 1 second. It then displays the currently selected thermocouple type
or mV for approximately 3 seconds. If °C/°F operation has been selected the
currently selected temperature scale of °C or °F will display.




![[ALTEK] 써머커플온도교정기 322-1 (단종) -> 대체모델 PIE 322-1 [ALTEK] 써머커플온도교정기 322-1 (단종) -> 대체모델 PIE 322-1](https://cdn-saas-web-116-121.cdn-nhncommerce.com/naraetek39_godomall_com/data/goods/13/04/10/115/1369061308550m0_detail.jpg)